PRODUCT

ELITex Staph

Latex slide agglutination test for the rapid identification of Staphylococcus aureus.

ELITex Staph is a slide agglutination test for the rapid identification of Staphylococcus aureus (MRSA included) from primary culture plates.

  • MRSA and MSSA identification
  • Cost-effective routine laboratory test
  • Easy and ready-to-use
  • High sensitivity and specificity
  • Fast results for clinicians (<1min)
  • Simple visual reading, without expert interpretation

The ELITex Staph assay is important in the management of patients with septicemia and nosocomial infections.

 

Reference Name Quantity
22711 / 22712 ELITex Staph 60 / 160 tests

 

ELITex Staph detection

Benefits

Intended use

Staphylococci are a serious problem for hospital epidemiology. Among them, Staphylococcus aureus is the main pathogenic species and is responsible for septicemia and nosocomial infections. Multi-drug resistance poses considerable therapeutic problems, particularly with respect to methicillin resistant staphylococci. The ability to rapidly differentiate Staphylococcus aureus from other staphylococci species is therefore a necessity in microbiological diagnosis. The ELITex Staph is a slide agglutination test for the rapid identification of Staphylococcus aureus (MRSA included) from primary culture plates. It is rapid, easy-to-use and easy-to-read with results available within 1 minute.


Principle

Red latex particles are sensitized with three proteins:

  • Fibrinogen that reacts with “Clumping Factor”.
  • The Fc fragment of human IgG that reacts with Staphylococcus aureus protein A.
  • Anti-Staphylococcus aureus capsular polysaccharides polyclonal IgG.

The presence of Staphylococcus aureus is revealed by a strong agglutination reaction. In the absence of Staphylococcus aureus, no agglutination appears. The test procedure is easy with results available in less than a minute. The kit is complete and includes:

  • TEST LATEX
  • CONTROL LATEX (internal control)
  • Test cards
  • Single use sticks

Simple protocol

 width=


Easy-to-read and easy-to-interpret results

TEST LATEX CONTROL LATEX Interpretation

Negative reaction

The strain tested is not Staphylococcus aureus

+
+

Positive reaction

The strain tested is Staphylococcus aureus

+ + Uninterpreted reaction

 width=

Reagents and material

Reference Quantity Description
22711 1 TEST LATEX: Dispenser vial of 1.8 mL of sensitized latex
1 CONTROL LATEX: Dispenser vial of 1.8 mL of unsensitized latex
15 TEST CARD: Disposable reaction card
120 STICK: Disposable mixing sticks
Reference Quantity Description
22712 2 TEST LATEX: Dispenser vial of 2.4 mL of sensitized latex
2 CONTROL LATEX: Dispenser vial of 2.4 mL of unsensitized latex
40 TEST CARD: Disposable reaction card
320 STICK: Disposable mixing sticks

Stability and storage

  • The reagents are ready-to-use.
  • All the reagents are stored at 2-8°C. Do not freeze.
  • The results of the stability study indicate that the product has a tability of 18 months at 2-8°C from its manufacture.

Performance

  • ELITex Staph assay has a good sensitivity for methicillin resistant strains of Staphylococcus aureus (MRSA), producing little protein A and/or fibrinogen affinity factor. The sensitivity of the assay is 99.02% and its specificity is 99.07%.

Material required but not supplied

  • Incubator calibrated at 37°C
  • Contaminated waste container

 

Manufacturer: ELITech Microbio.
Product(s) intended for healthcare professionals.
Read the instructions on the label and/or instructions for use of the product(s).